set_output_limit

Set the output voltage limits of a given channel

Parameters

integer required
Target channel
allowed values:
Moku:Lab1, 2
Moku:Go1, 2
Moku:Pro1, 2, 3, 4
Moku:Delta1, 2, 3, 4
boolean required
Enable the output voltage limiter
allowed values:true, false
number
Lower voltage limit of the output, used when the limiter is enabled
default: None
allowed values:
Moku:Lab-1 to 1
Moku:Go-5 to 5
Moku:Pro-1 to 1
Moku:Delta-0.5 to 0.5
units: V
number
Upper voltage limit of the output, used when the limiter is enabled
default: None
allowed values:
Moku:Lab-1 to 1
Moku:Go-5 to 5
Moku:Pro-1 to 1
Moku:Delta-0.5 to 0.5
units: V
boolean
Disable all implicit conversions and coercions
default: true
allowed values:true, false

# Examples

# Sample response

{
    "enable": true,
    "high_limit": 0.5,
    "low_limit": -0.5
}
1
2
3
4
5

When enable is false, only the enable field is returned.

{
    "enable": false
}
1
2
3