Sets Phasemeter loop parameters

Configures the frequency and bandwidth of the PLL for each channel

Parameters

integer required
Target channel
allowed values:
Moku:Lab1, 2
Moku:Go1, 2
Moku:Pro1, 2, 3, 4
boolean
Automatically acquire the initial frequency of the input signal
default: false
number
Frequency of the input signal
default: 1000000
allowed values:
Moku:Lab1e3 to 200e6
Moku:Go1e3 to 30e6
Moku:Pro1e3 to 300e6
units: Hz
number
Bandwidth of the phase-locked loop
default: 1kHz
allowed values:
Moku:Lab1Hz, 10Hz, 100Hz, 1kHz, 10kHz, 100kHz
Moku:Go1Hz, 10Hz, 100Hz, 1kHz, 10kHz, 100kHz
Moku:Pro1Hz, 10Hz, 100Hz, 1kHz, 10kHz, 100kHz, 1MHz
units: Hz
boolean
Disable all implicit conversions and coercions.
default: true

# Examples